WJ303A power analyzer
1. Measurement accuracy index (only measure 45~65Hz power frequency range signal)
Conditions that meet the error requirements
• Temperature: (23±5)°C
• Humidity: (30~75)%RH
• Input waveform: Sine wave
• Common mode voltage: 0V
• Magnification Setting: Set to OFF
• Preheat for 30 minutes
parameter |
Measuring range |
error |
Voltage U |
9.0~1000.0V |
Signal under test: 45 Hz ≤ f ≤ 65 Hz, When the signal under test is < (range*10%), error: ± (0.40% of reading + 0.10% of range) When the measured signal is ≥ (range*10%), error: ± (0.16% of reading + 0.04% of range) |
Current I |
0.020~5.000A |
|
Active power P |
Voltage range * current range |
Signal under test: 45 Hz ≤ f ≤ 65 Hz, at rated voltage 380 V, Error when power factor PF=1.0: ± (0.16% of reading + 0.04% of range) When power factor PF=0.5, error: ± (0.80% of reading + 0.20% of range) |
Power factor PF |
0.05~1.00 |
Signal under test: 45 Hz ≤ f ≤ 65 Hz, error: ±0.01 |
Voltage frequency UF
|
45.00-65.00Hz |
Counting method, the voltage (or current) value is U (or I), and the voltage (or current) range is UR (or IR). When U (or I) ≥ (UR (or IR) * 10%), the error is 0.1% * reading; When U (or I) ≥ (UR (or IR) * 5%) and < (UR (or IR) * 10%), the error is 0.2% * reading; When U (or I) < (UR (or IR) * 5%), the frequency data is only for reference because the signal is too small. |
Current frequency IF |
45.00-65.00Hz |
Gauge specifications
The performance indicators of the tester in numerical form are within the specified tolerances. Within 30 minutes of power-on and 2 full data acquisition cycles, the tester met the specified accuracy. Performance metrics are based on a one-year calibration cycle.
Gauge range
|
Conventional range |
Other special ranges |
Voltage range UN |
9.0V~1000.0V (45--65Hz AC voltage signal) |
to the product on The range indicator shall prevail |
Current range IN |
0.02A~5.00A (45--65Hz AC current signal) |
Other technical indicators
• Signal input and impedance: The voltage and current are all floating inputs, with a voltage input impedance of about 1 MΩ and a current input impedance of about 4mΩ.
• Maximum peak value of measurement signal: 1.6 times of the maximum range of voltage and current.
• Sampling rate: 8000 times/s, sampling period of about 125μS, 24 bits, voltage and current sampling at the same time.
• Zero calibration mode: automatic zero calibration when the machine is turned on
• Line filtering: fixed at 2kHz
• Data update: The data calculation period can be set from 0.05 seconds ~ 1 second.
• Display: 7 windows LED display.
• Power consumption: <10VA.
• Gauge weight: approx. 3.5 kg.
• Maximum gauge dimensions: width× height×depth (223.5×151.5×384).
• Gauge cut-out size: width × height (209x128).
Safety requirements
• Insulation resistance: The insulation resistance between the measuring terminal and the instrument housing is not less than 5MΩ;
• Voltage resistance: 2000V 50Hz sine wave voltage between the measuring terminal and the instrument shell;
Working environment
• 温度:(0~40)℃;
• 湿度:(20%~75%)RH;
• 大气压力: (86~106)kPa;
• 仪表工作电源:AC 100V~240V 50/60Hz 无较重的振动及电磁干扰;
• 以上技术参数的说明中所用到的术语定义请参见GB/T 13978-2008 《数字多用表通用技术条件》。
Recommended Products
Jinyuan wholeheartedly welcomes new and old customers to visit and guide, and sincerely hopes to share technology with you and create the future together!
CONTACT US
Telephone:027-87051278
Fax:027-87051228
Mail:whjydq_666@163.com
Address: Building 5, Donghu High tech Industrial Innovation Base, No. 9 Miaoshan Avenue, Jiangxia District, Wuhan City
Official public account

Mobile website

Copyright © Wuhan Jinyuan Electric Co., Ltd. All rights reserved. ICP:NO.17006198-1
This site supports IPv6 access | SEO